BSD-2000 Deep Phased Array
Commercial Product Comparative Table
BSD-2000B | BSD-2000-3D | BSD-2000-3D/MR | |
---|---|---|---|
RF Generator | Tetra | Dodek | Dodek |
# of RF Channels | 4 | 12 | 12 |
Directional Heat Zone Steering | 2 axis (X & Y) | 3 axis (X, Y & Z) | 3 axis (X, Y & Z) |
Maximum Power | 1300 | 1800 | 1800 |
Max. Channel Power | 325 | 450 (4) or 150 (12) | 150 |
Frequency (MHz) | 75 to 140 MHz | 75 to 140 MHz | 100 MHz |
# of Antennas | 8 | 8 and 24 | 24 |
Forward Pwr Cont. | Yes | Yes | Yes |
Phase Control | Yes | Yes | Yes |
MRI Compatible | No | No | Yes |
Multiple size applicators | Yes | Yes | Yes |
Computer Console | included | included | included |
Sigma 30 (30cm) applicator | Yes | Yes | No |
Sigma 60 (60cm) applicator | Yes | Yes | No |
Sigma Ellipse applicator | Yes | Yes | No |
Sigma Eye applicator | No | Yes | No |
Sigma 30MR applicator | No | No | Yes |
Sigma Eye MR applicator | No | No | Yes |
# Temperature probes | 8 | 8 | 8 |
# Sensors per probe | 1 | 1 | 1 |
# of Sensor Locations Along Catheter | 31 | 31 | 31 |
Thermal Mapping | Yes | Yes | Yes |
Built-in Thermal Calibration | Yes | Yes | Yes |
CE Mark | Yes | Yes | Yes |
FDA Clearance in U.S. | Yes HDE | in process | in process |
Forward Pwr. Detection | Yes | Yes | Yes |
Reflected Pwr. Detection | Yes | Yes | Yes |
Water Heating/Cooling System | Built-in | Built-in | External |
Water Bolus Interface | Yes | Yes | Yes |
Variable Focal Diameter | Yes | Yes | Yes |
Non-invasive thermometry | No | No | Yes |
Published Clinical Trials | Yes | Yes | Yes |
# of Water Control Systems Needed | 1 | 1 | 1 |
ESHO Guideline Compliant | Yes | Yes | Yes |
Automatic Bolus Air Purge | Yes | Yes | Yes |
UV Water Treatment | Yes | Yes | Yes |
Integral Deionizer of Water | Yes | Yes | Yes |
Currently Available For Sale | Yes | Yes | Yes |
Pretreatment Planning | Yes | Yes | Yes |
Shield Room Required | Yes | Yes | Uses Existing MRI Shielding |